TU Darmstadt / ULB / TUprints

On the performance of wavelength meters: Part 2 — frequency-comb based characterization for more accurate absolute wavelength determinations

König, Kristian ; Imgram, Phillip ; Krämer, Jörg ; Maaß, Bernhard ; Mohr, Konstantin ; Ratajczyk, Tim ; Sommer, Felix ; Nörtershäuser, Wilfried (2024)
On the performance of wavelength meters: Part 2 — frequency-comb based characterization for more accurate absolute wavelength determinations.
In: Applied Physics B : Lasers and Optics, 2020, 126 (5)
doi: 10.26083/tuprints-00023873
Article, Secondary publication, Publisher's Version

[img] Text
s00340-020-07433-4.pdf
Copyright Information: CC BY 4.0 International - Creative Commons, Attribution.

Download (1MB)
Item Type: Article
Type of entry: Secondary publication
Title: On the performance of wavelength meters: Part 2 — frequency-comb based characterization for more accurate absolute wavelength determinations
Language: English
Date: 8 April 2024
Place of Publication: Darmstadt
Year of primary publication: May 2020
Place of primary publication: Berlin ; Heidelberg
Publisher: Springer
Journal or Publication Title: Applied Physics B : Lasers and Optics
Volume of the journal: 126
Issue Number: 5
Collation: 8 Seiten
DOI: 10.26083/tuprints-00023873
Corresponding Links:
Origin: Secondary publication DeepGreen
Abstract:

Wavelength meters are widely used for frequency determinations and stabilization purposes since they cover a large wavelength range, provide a high read-out rate and have specified accuracies of up to 10⁻⁸. More accurate optical frequency measurements can be achieved with frequency combs but only at the price of considerably higher costs and complexity. In the context of precise and accurate frequency determinations for high-resolution laser spectroscopy, the performance of five different wavelength meters was quantified with respect to a frequency comb. The relative precision as well as the absolute accuracy has been investigated in detail, allowing us to give a sophisticated uncertainty margin for the individual instruments. We encountered a prominent substructure on the deviation between both device types with an amplitude of a few MHz that is repeating on the GHz scale. This finally limits the precision of laser scans which are monitored and controlled with wavelength meters. While quantifying its uncertainty margins, we found a high temporal stability in the characteristics of the wavelength meters which enables the preparation of wavelength-dependent adjustment curves for wide- and short-ranged scans. With this method, the absolute accuracy of wavelength meters can be raised up to the MHz level independently from the wavelength of the reference laser used for calibrating the device. Since this technique can be universally applied, it can lead to benefits in all fields of wavelength meter applications.

Uncontrolled Keywords: Physics, general, Physical Chemistry, Optics, Lasers, Photonics, Optical Devices, Quantum Optics, Engineering, general
Identification Number: Artikel-ID: 86
Status: Publisher's Version
URN: urn:nbn:de:tuda-tuprints-238739
Classification DDC: 500 Science and mathematics > 530 Physics
Divisions: 05 Department of Physics > Institute of Nuclear Physics
Date Deposited: 08 Apr 2024 12:42
Last Modified: 10 Apr 2024 05:59
SWORD Depositor: Deep Green
URI: https://tuprints.ulb.tu-darmstadt.de/id/eprint/23873
PPN: 516998293
Export:
Actions (login required)
View Item View Item