TU Darmstadt / ULB / TUprints

Towards an Optical Gas Standard for Traceable Calibration-Free and Direct NO₂ Concentration Measurements

Nwaboh, Javis A. ; Qu, Zhechao ; Werhahn, Olav ; Ebert, Volker (2024)
Towards an Optical Gas Standard for Traceable Calibration-Free and Direct NO₂ Concentration Measurements.
In: Applied Sciences, 2021, 11 (12)
doi: 10.26083/tuprints-00019526
Article, Secondary publication, Publisher's Version

[img]
Preview
Text
applsci-11-05361-v2.pdf
Copyright Information: CC BY 4.0 International - Creative Commons, Attribution.

Download (1MB) | Preview
Item Type: Article
Type of entry: Secondary publication
Title: Towards an Optical Gas Standard for Traceable Calibration-Free and Direct NO₂ Concentration Measurements
Language: English
Date: 15 January 2024
Place of Publication: Darmstadt
Year of primary publication: 2021
Place of primary publication: Basel
Publisher: MDPI
Journal or Publication Title: Applied Sciences
Volume of the journal: 11
Issue Number: 12
Collation: 10 Seiten
DOI: 10.26083/tuprints-00019526
Corresponding Links:
Origin: Secondary publication DeepGreen
Abstract:

We report a direct tunable diode laser absorption spectroscopy (dTDLAS) instrument developed for NO₂ concentration measurements without chemical pre-conversion, operated as an Optical Gas Standard (OGS). An OGS is a dTDLAS instrument that can deliver gas species amount fractions (concentrations), without any previous or routine calibration, which are directly traceable to the international system of units (SI). Here, we report NO₂ amount fraction quantification in the range of 100–1000 µmol/mol to demonstrate the current capability of the instrument as an OGS for car exhaust gas application. Nitrogen dioxide amount fraction results delivered by the instrument are in good agreement with certified values of reference gas mixtures, validating the capability of the dTDLAS-OGS for calibration-free NO₂ measurements. As opposed to the standard reference method (SRM) based on chemiluminescence detection (CLD) where NO₂ is indirectly measured after conversion to NO, titration with O₃ and the detection of the resulting fluorescence, a dTDLAS-OGS instrument has the benefit of directly measuring NO₂ without distorting or delaying conversion processes. Therefore, it complements the SRM and can perform fast and traceable measurements, and side-by-side calibrations of other NO₂ gas analyzers operating in the field. The relative standard uncertainty of the NO₂ results reported in this paper is 5.1% (k = 1, which is dominated (98%) by the NO₂ line strength), the repeatability of the results at 982.6 µmol/mol is 0.1%, the response time of the instrument is 0.5 s, and the detection limit is 825 nmol/mol at a time resolution of 86 s.

Uncontrolled Keywords: gas analysis, laser absorption spectroscopy, dTDLAS, nitrogen dioxide (NO₂), metrology
Status: Publisher's Version
URN: urn:nbn:de:tuda-tuprints-195260
Additional Information:

This article belongs to the Special Issue Diode Laser Spectroscopy – Robust Sensing for Environmental and Industrial Applications

Classification DDC: 500 Science and mathematics > 530 Physics
600 Technology, medicine, applied sciences > 620 Engineering and machine engineering
600 Technology, medicine, applied sciences > 660 Chemical engineering
Divisions: 16 Department of Mechanical Engineering > Institute of Reactive Flows and Diagnostics (RSM)
Date Deposited: 15 Jan 2024 13:35
Last Modified: 13 Mar 2024 07:04
SWORD Depositor: Deep Green
URI: https://tuprints.ulb.tu-darmstadt.de/id/eprint/19526
PPN: 516179268
Export:
Actions (login required)
View Item View Item