Dosimetric Validation of a System to Treat Moving Tumors Using Scanned Ion Beams That Are Synchronized With Anatomical Motion

Lis, Michelle ; Newhauser, Wayne ; Donetti, Marco ; Wolf, Moritz Ernst ; Steinsberger, Timo ; Paz, Athena ; Graeff, Christian (2023)
Dosimetric Validation of a System to Treat Moving Tumors Using Scanned Ion Beams That Are Synchronized With Anatomical Motion.
In: Frontiers in Oncology, 2021, 11
doi: 10.26083/tuprints-00019984
Article, Secondary publication, Publisher's Version

[img]
Preview
Text
fonc-11-712126.pdf
Copyright Information: CC BY 4.0 International - Creative Commons, Attribution.

Download (1MB) | Preview
Item Type: Article
Type of entry: Secondary publication
Title: Dosimetric Validation of a System to Treat Moving Tumors Using Scanned Ion Beams That Are Synchronized With Anatomical Motion
Language: English
Date: 14 November 2023
Place of Publication: Darmstadt
Year of primary publication: 2021
Place of primary publication: Lausanne
Publisher: Frontiers Media S.A.
Journal or Publication Title: Frontiers in Oncology
Volume of the journal: 11
Collation: 13 Seiten
DOI: 10.26083/tuprints-00019984
Corresponding Links:
Origin: Secondary publication DeepGreen
Abstract:

Purpose: The purpose of this study was to validate the dosimetric performance of scanned ion beam deliveries with motion-synchronization to heterogenous targets.

Methods: A 4D library of treatment plans, comprised of up to 10 3D sub-plans, was created with robust and conventional 4D optimization methods. Each sub-plan corresponded to one phase of periodic target motion. The plan libraries were delivered to a test phantom, comprising plastic slabs, dosimeters, and heterogenous phantoms. This phantom emulated range changes that occur when treating moving tumors. Similar treatment plans, but without motion synchronization, were also delivered to a test phantom with a stationary target and to a moving target; these were used to assess how the target motion degrades the quality of dose distributions and the extent to which motion synchronization can improve dosimetric quality. The accuracy of calculated dose distributions was verified by comparison with corresponding measurements. Comparisons utilized the gamma index analysis method. Plan quality was assessed based on conformity, dose coverage, overdose, and homogeneity values, each extracted from calculated dose distributions.

Results: High pass rates for the gamma index analysis confirmed that the methods used to calculate and reconstruct dose distributions were sufficiently accurate for the purposes of this study. Calculated and reconstructed dose distributions revealed that the motion-synchronized and static deliveries exhibited similar quality in terms of dose coverage, overdose, and homogeneity for all deliveries considered. Motion-synchronization substantially improved conformity in deliveries with moving targets. Importantly, measurements at multiple locations within the target also confirmed that the motion-synchronized delivery system satisfactorily compensated for changes in beam range caused by the phantom motion. Specifically, the overall planning and delivery approach achieved the desired dose distribution by avoiding range undershoots and overshoots caused by tumor motion.

Conclusions: We validated a dose delivery system that synchronizes the movement of the ion beam to that of a moving target in a test phantom. Measured and calculated dose distributions revealed that this system satisfactorily compensated for target motion in the presence of beam range changes due to target motion. The implication of this finding is that the prototype system is suitable for additional preclinical research studies, such as irregular anatomic motion.

Uncontrolled Keywords: motion-synchronized dose delivery, carbon ion therapy, range compensation, motion mitigation, multi-phase 4D delivery, scanned ion beams
Status: Publisher's Version
URN: urn:nbn:de:tuda-tuprints-199844
Classification DDC: 500 Science and mathematics > 530 Physics
600 Technology, medicine, applied sciences > 610 Medicine and health
Divisions: 05 Department of Physics > Institute for Condensed Matter Physics
Date Deposited: 14 Nov 2023 14:16
Last Modified: 16 Nov 2023 07:03
SWORD Depositor: Deep Green
URI: https://tuprints.ulb.tu-darmstadt.de/id/eprint/19984
PPN: 51321920X
Export:
Actions (login required)
View Item View Item